Configuración de xDrip+

If not already set up then download xDrip+.

This documentation is for xDrip+ for Android only. There is an app “xDrip for iOS” that has nothing to do with the original xDrip+ for Android.

Para los transmisores G6 fabricados después de otoño/finales de 2018 (es decir, con números de serie starting with 80 or 81) you can use the master version.

Si el número de serie de su transmisor Dexcom G6 is starting with 8G..., 8H... or 8J... use one of the latest nightly builds.

If your phone runs Android 10 and you have difficulties with xDrip+ master try nightly build 2019/12/31 or later.

Configuración básica para todos los medidores contínuos (CGM), sistemas CGM puros o sistemas flash (FGM)

  • Asegúrate de establecer la URL base correctamente utilizando https:// y no http://

    Por ejemplo, https://TU_API_SECRET@nombre_de_tu_app.herokuapp.com/api/v1/

    -> Pulsa en la hamburguesa (menú con tres líneas de la parte superior izquierda de la pantalla) -> Configuración-> Subir a la nube-> Sincronización con Nightscout (REST-API) -> URL base

  • Desactiva Calibración automática Si la caja de verificación Automatic Calibration está marcada, activa Descargar datos y a continuación desmarca la caja de verificación de Calibración automática y desactiva Descargar datos de nuevo, de lo contrario los tratamientos (insulina y carbohidratos) se añadirán dos veces a Nightscout.

  • Pulsa Opciones adicionales

  • Desactiva Cargar tratamientos y Rellenar datos.

    Safety warning : You must deactivate “Upload treatments” from xDrip+, otherwise treatments can be doubled in AAPS leading to false COB and IOB.

  • La opción Alerta de fallos también debe desactivarse. De lo contrario, recibirás una alarma cada 5 minutos si la red wifi/móvil es demasiado mala o si el servidor no está disponible.

    xDrip+ Ajustes Básicos 1

    xDrip+ Ajustes Básicos 2

  • Configuración InterApps Si vas a utilizar AndroidAPS es necesario activar <0>Difusión local</0> dentro de las opciones InterApps de xDrip+ para que los datos sean emitidos y AndroidAPS pueda leerlos.

  • Para que los valores sean consistentes, debes activar Enviar el valor de glucosa visualizado.

  • Si has activado Aceptar tratamientos en xDrip+ y <0>Difusión de datos</0> en AndroidAPS, entonces xDrip+ recibirá los datos de insulina, carbohidratos y dosis basal desde AndroidAPS para poder realizar predicciones de hipoglucemia, etc. con mayor precisión.

    xDrip+ Ajustes Básicos 3

Identificar receptor

  • Si descubres problemas con la difusión local (AAPS no recibe los valores de glucemia en sangre (BG) de xDrip+) ve a Ajustes > Ajustes InterApp > Identificar receptor y especifica info.nightscout.androidaps.

  • Importante: la corrección automática a veces tiende a cambiar la “i” de “info” por una mayúscula. Debes utilizar solo letras minúsculas al escribir info.nightscout.androidaps. Capital I would prevent AAPS from receiving BG values from xDrip+.

    xDrip+ Inter-app Ajustes Básicos Identificar receptor

xDrip+ & Dexcom G6

  • El transmisor Dexcom G6 puede conectarse simultáneamente al receptor Dexcom (o alternativamente a la bomba t:slim) y a una aplicación en tu teléfono.
  • Para usar xDrip+ como receptor, desinstala primero la aplicación Dexcom. No se puede conectar al mismo tiempo la aplicación xDrip+ y Dexcom con el transmisor.
  • Si necesitas utilizar Clarity (app Dexcom) y quieres beneficiarte de las alarmas de xDrip+ utiliza la aplicación Dexcom parcheada y la opción de difusión local a xDrip+.

La versión de xDrip+ a utilizar depende del número de serie de los transmisores G6.

  • Para los transmisores G6 fabricados después de otoño/finales de 2018 (es decir, con números de serie starting with 80 or 81) you can use the master.
  • Si el número de serie de su transmisor Dexcom G6 is starting with 8G, 8H or 8J try nightly build 2019/07/28 or later.

Ajustes específicos de Dexcom

  • Abra G5/G6 Configuración de depuración -> Menú Hamburguesa (parte superior izquierda de la pantalla de inicio) -> Settings -> G5/G6 Debug Settings Abrir los ajustes de xDrip+

  • Habilite los siguientes parámetros

    • Usar el Colector OB1
    • Algoritmo nativo (importante si desea utilizar SMB)
    • Soporte G6
    • Permitir desvinculación de OB1
    • Permitir a OB1 iniciar la vinculación
  • Todas las demás opciones deben ser desactivados

  • Ajuste el nivel de aviso de la batería a 280 (parte inferior de los valores de depuración G5/G6)

    ajustes de depuración de xDrip+ G5/G6

No se recomiendan reinicios preventivos

With Dexcom transmitters who’s serial no. is starting with 8G, 8H or 8J preemptive restarts do not work and might kill the sensor completely!

The automatic extension of Dexcom sensors (preemptive restarts) is not recommended as this might lead to “jumps” in BG values on day 9 after restart.

xDrip+ Jump after Preemptive Restart

Lo que está claro es que el uso del G6 es tal vez un poco más complejo que lo que se sugiere en primer lugar. Para utilizar de manera segura, hay un par de puntos a tener en cuenta:

  • Si estás usando los datos nativos con el código de calibración en xDrip+ o Spike, la cosa más segura es no permitir reinicios preventivos del sensor.
  • Si debe usar reinicio preventivo, a continuación, asegúrese de insertar una hora del día donde puede observar el cambio y calibrar si es necesario.
  • Si está reiniciando sensores, o bien haga esto sin la calibración de fábrica para obtener resultados más seguros en los días 11 y 12, o asegúrate de que estás listo para calibrar y vigilar la variación.
  • La preabsorción del G6 con calibración de fábrica es probable que dé variaciones en los resultados. Si haces preconfiguración, entonces para obtener los mejores resultados, probablemente necesitarás calibrar el sensor.
  • Si no estás observando los cambios que pueden estar teniendo lugar, puede ser mejor volver al modo no calibrado en fábrica y usar el sistema como un G5.

To learn more about the details and reasons for these recommendations read the complete article published by Tim Street at www.diabettech.com.

Conectar el transmisor G6 por primera vez

For second and following transmitters see Extend transmitter life below.

Para los transmisores G6 fabricados después de otoño/finales de 2018 (es decir, con números de serie starting with 80 or 81) you can use the master.

Si el número de serie de su transmisor Dexcom G6 is starting with 8G, 8H or 8J try nightly build 2019/07/28 or later.

  • Desactive el receptor Dexcom original (si se utiliza).

  • Pulse el icono xDrip+ en la gota de sangre de color rojo en la pantalla principal para habilitar el botón Asistente de Origen.

  • Utilice el Botón Asistente de Origen que asegura la configuración predeterminada incluyendo OB1 & Modo Nativo

    • Esto le guiará a través de la configuración inicial.
    • necesitará el número de serie del transmisor si es la primera vez que lo utiliza.
  • Ponga el número de serie del nuevo transmisor (en el envase del transmisor o en la parte posterior del transmisor). Tenga cuidado de no confundir 0 (cero) y O (letra o mayúscula).

    xDrip+ Dexcom número de serie del Transmisor

  • Insertar nuevo sensor (sólo si se sustituye)

  • Coloque el transmisor en el sensor

  • No inicie un nuevo sensor antes de que se muestre la información siguiente en la página de estado clásica-> G5/G6 estado-> PhoneServiceState:

    • Transmisor con número de serie que empieza por 80 o 81: “Obtiene datos hh:mm” (por ejemplo, “Obtiene datos 19:04”)
    • Transmitter serial starting with 8G, 8H or 8J: “Got glucose hh:mm” (i.e. “Got glucose 19:04”) or “Got no raw hh:mm” (i.e. “Got now raw 19:04”)

    xDrip+ PhoneServiceState

  • Iniciar sensor (sólo si se reemplaza)

    -> Cerca de la parte inferior de la pantalla Iniciando x, x horas faltantes debe aparecer después de unos pocos minutos.

-> If your transmitter serial no. does not start with 8G, 8H or 8J and there is no time specification after a few minutes stop and restart the sensor.

  • Reiniciar el colector (estado del sistema - si no se reemplaza el sensor}

  • No encienda el receptor original de Dexcom (si lo utiliza) de nuevo antes de xDrip+ muestre las primeras lecturas.

  • Pulse el icono xDrip+ de goteo de sangre de color rojo en la pantalla principal para inhabilitar el Botón Asistente de origen.

    xDrip+ Dexcom Transmisor 1

    xDrip+ Dexcom Transmisor 2

    xDrip+ Dexcom Transmisor 3

    xDrip+ Dexcom Transmisor 4

Estado de la batería de Transmisor

  • El estado de la batería se puede controlar en el estado del sistema (menú Hamburguesa en la parte superior izquierda de la pantalla de inicio)
  • Deslizar la izquierda una vez para ver la segunda pantalla. xDrip+ First Transmisor 4
  • No se conocen los valores exactos cuando el transmisor “muere” debido a una batería vacía. La información siguiente se publicó en línea después de que el transmisor “murió”:
    • Publicación 1: Día de transmisión: 151 / Voltaje A: 297 / Voltaje B: 260 / Resistencia: 2391
    • Publicación 2: Días de transmisión: 249 / Voltaje A: 275 (en tiempo de falla)

Extender vida del transmisor

  • So far life cannot be extended for transmitters who’s serial no. starts with 8G, 8H or 8J. Same is true for transmitters with serial no. starting with 81 and firmware 1.6.5.27 (see xDrip+ System Status - G5/G6 status as shown in screenshot above).

  • Para evitar las dificultades iniciando los sensores, se recomienda encarecidamente extender la vida del transmisor antes del día 100 de la primera utilización.

  • Use of transmitters serial no. starting with 81 and firmware 1.6.5.27 beyond day 100 is only possible if ‘engineering mode’ is turned on and ‘native mode’ is deactivated (hamburger menu -> settings -> G5/G6 debug settings -> native algorithm) because a transmitter hard reset is NOT possible.

  • Running sensor session will be stopped when extending transmitter life. So, extend before sensor change or be aware that there will be a new 2 h warm-up phase.

  • Stop sensor manually via hamburger menu.

  • Switch to the engineering mode:

    • tap on the character on the right of the xDrip+ start screen that represents a syringe
    • then tap on the microphone icon in the lower right corner
    • In the text box that opens type “enable engineering mode”
    • click “Done”
    • If Google Speak engine is enabled, you can also speak the voice command: “enable engineering mode”.
  • Go to the G5 debug settings and make sure Use the OB1 collector is enabled.

  • Use the voice command: “hard reset transmitter”

  • The voice command will be executed with the next data receipt of the transmitter

  • Look at the system status (Hamburger menu -> system status) and see what happens

  • After approx. 10 min. you can switch to ‘Classic Status Page’ (swipe right) and click ‘Restart collector’. This will set sensor days to 0 without the need to start a new sensor.

  • Alternative: If you see a message “Phone Service State: Hard Reset maybe failed” on second system status screen just start the sensor and this message should go away.

    xDrip+ Hard Reset maybe failed

  • Transmitter days will be set to 0 after successful extension and start of sensor.

Sustituir transmisor

Para los transmisores G6 fabricados después de otoño/finales de 2018 (es decir, con números de serie starting with 80 or 81) you can use the master.

Si el número de serie de su transmisor Dexcom G6 is starting with 8G, 8H or 8Juse one of the latest nightly builds.

  • Desactive el receptor Dexcom original (si se utiliza).

  • Detener sensor (sólo si sustituye el sensor)

    Asegúrese de que realmente se ha detenido:

    En la segunda pantalla “Estado de G5/G6”, mire a Elementos de cola aproximadamente en la mitad hacia abajo - Puede que diga algo como (1) Detener sensor

    Espere hasta que esto ocurra - normalmente en unos pocos minutos. El estado del sensor debe ser “Detenido” (consulte la captura de pantalla).

    -> Para quitar el transmisor sin detener el sensor ver este video https://youtu.be/AAhBVsc6NZo.

    xDrip+ Detener sensor Dexcom 1

    xDrip+ Detener sensor Dexcom 2

  • Forget device in xDrip+ system status AND in smartphone’s BT settings (Will be shown as Dexcom?? mientras ?? son los dos últimos dígitos del número de serie del transmisor)

    xDrip+ Olvidar dispositivo

  • Retire el transmisor (y el sensor si reemplaza el sensor)

  • Coloque el transmisor antiguo lejos para evitar la reconexión. Un microondas es un escudo de Faraday perfecto para esto, pero desenchufa el cable de alimentación para estar seguro al 100% de que nadie lo encienda.

  • Pulse el icono xDrip+ en la gota de sangre de color rojo en la pantalla principal para habilitar el botón Asistente de Origen.

  • Utilice el Botón Asistente de Origen que asegura la configuración predeterminada incluyendo OB1 & Modo Nativo

    • Esto le guiará a través de la configuración inicial.
    • Necesitará el número de serie del transmisor si es la primera vez que lo utiliza.
  • Introducir el número de serie del transmisor nuevo. Tenga cuidado de no confundir 0 (cero) y O (letra o mayúscula).

  • Insertar nuevo sensor (sólo si se sustituye).

  • Coloque el transmisor en el sensor - No iniciar el sensor inmediatamente!

  • Nuevos “Transmisores Firefly” (número de serie starting with 8G, 8H or 8J) can only be used in native mode.

  • Las opciones siguientes no deben activarse para los nuevos “Transmisores Firefly” (número de serie starting with 8G, 8H or 8J):

    • Reinicio preventivo (inhabilitar)
    • Reiniciar el sensor (deshabilitar!)
    • Fallback to xDrip+ (disable!)

    Configuración de los transmisores Firefly

  • Check in Classic Status Page -> G5/G6 status -> PhoneServiceState if one of the following information is displayed:

    • Transmisor con número de serie que empieza por 80 o 81: “Obtiene datos hh:mm” (por ejemplo, “Obtiene datos 19:04”)
    • Transmitter serial starting with 8G, 8H or 8J: “Got glucose hh:mm” (i.e. “Got glucose 19:04”) or “Got no raw hh:mm” (i.e. “Got now raw 19:04”)

    xDrip+ PhoneServiceState

  • Espere 15 minutos ya que el transmisor debe comunicarse varias veces con xDrip antes de que se inicie el nuevo sensor. La batería de datos se muestra a continuación la información del Firmware.

    Datos de batería del transmisor firefly

  • ¡Inicie el sensor y NO PREDATE LA FECHA! Seleccione siempre “Sí, hoy”!

  • Reiniciar el colector (estado del sistema - si no se reemplaza el sensor)

  • No encienda el receptor original de Dexcom (si lo utiliza) de nuevo antes de xDrip+ muestre las primeras lecturas.

  • Pulse el icono xDrip+ de goteo de sangre de color rojo en la pantalla principal para inhabilitar el Botón Asistente de origen.

    xDrip+ Dexcom Transmisor 1

    xDrip+ Dexcom Transmisor 2

    xDrip+ Dexcom Transmisor 3

    xDrip+ Dexcom Transmisor 4

Nuevo Sensor

  • Desactive el receptor Dexcom original (si se utiliza).

  • Detener sensor si es necesario

    Asegúrese de que realmente se ha detenido:

    En la segunda pantalla “Estado de G5/G6”, mire a Elementos de cola aproximadamente en la mitad hacia abajo - Puede que diga algo como (1) Detener sensor

    Espere hasta que esto ocurra - normalmente en unos pocos minutos.

    xDrip+ Detener sensor Dexcom 1

    xDrip+ Detener sensor Dexcom 2

  • Limpiar los contactos (lado trasero del transmisor) con alcohol y dejemos secar al aire.

  • En caso de que utilice esta función, inhabilite Reiniciar el sensor y Reinicios anticipados (menú Hamburgesa-> Valores-> G5/G6 Depuración de valores). Si se pierda este paso y tiene estas funciones habilitadas, el nuevo sensor no se iniciará correctamente.

    xDrip+ Reinicio preventivo

  • Iniciar Sensor

    Para los nuevos transmisores Firefly (número de serie starting with 8G, 8H or 8J) it is mandatory, for all other transmitters it is recommended to wait approx. 15 minutes between stopping and starting the new sensor (until Sensor Status: Stopped is shown on second system status screen). NO PREDATE FECHAS!

  • Establecer tiempo insertado

    • Para utilizar el modo Nativo G6, debe esperar a que se inicie son 2 horas (por ejemplo, el tiempo de inserción es ahora).
    • Si utiliza el algoritmo xDrip+, puede establecer un tiempo anterior de más de 2 horas para evitar el inicio. Las lecturas pueden ser muy erráticas. Por lo tanto, esto no se recomienda.
  • Introduzca el código del Sensor (en la lámina de papel de aluminio del sensor)

    • Mantenga el código para futuras referencias (por ejemplo, un nuevo inicio después de que el transmisor haya tenido que eliminarse)
    • También se puede encontrar código en xDrip+ logs : pulse el menú de 3 puntos en la pantalla de inicio de xDrip+ y elija Ver registros de sucesos.
  • No se necesita ninguna calibración si utiliza G6 en “modalidad nativa”. xDrip+ mostrará las lecturas automáticamente después del inicio de 2 horas.

  • No encienda el receptor original de Dexcom (si lo utiliza) de nuevo, antes de que xDrip+ muestre las primeras lecturas.

    xDrip+ Inicio Dexcom Sensor 1

    xDrip+ Inicio Dexcom Sensor 2

Recuperar código de sensor

  • En el maestro fechado 2019/05/18 y las últimas Versiones el código del sensor se muestra en el estado del sistema (la parte superior del menú de Hamburgesa que queda en la pantalla de inicio).

  • Deslizar la izquierda una vez para ver la segunda pantalla.

    xDrip+ Recuperar código2 sensor Dexcom

  • El código de sensor de Dexcom también se puede encontrar en los registros de xDrip+.

  • Pulse el menú de 3 puntos (parte superior derecha en la pantalla de inicio)

  • Seleccione Ver registros de sucesos y busque “código”

    xDrip+ Recuperar código sensor Dexcom

Solución de problemas de Dexcom G5/G6 y xDrip+

Problema al conectar el transmisor

  • El transmisor debe mostrarse en la configuración bluetooth de su smartphone.
  • El transmisor se muestra como Dexcom?? mientras ?? representan los dos últimos dígitos del número de serie del transmisor (por ejemplo, DexcomHY).
  • Abrir el estado del sistema en xDrip+ (menú hamburguesa en la parte superior izquierda de la pantalla de inicio).
  • Compruebe si el transmisor se muestra en la primera página de estado (‘página de estado clásico’).
  • Si no, elimine el dispositivo de la configuración bluetooth de su smartphone y reinicie el recolector.
  • Espere unos 5 minutos. hasta que el transmisor Dexcom se vuelva a conectar automáticamente.

Problema al iniciar un nuevo sensor

Please note that the following method might likely not work if your Dexcom G6 transmitter’s serial no. is starting with 8G, 8H or 8J.

  • El sensor nativo se marca como “FAILED: Sensor Failed Start”
  • Detener Sensor
  • Reinicia tu dispositivo
  • Iniciar el sensor con el código 0000 (cuatro veces cero)
  • Espere 15 minutos
  • Detener Sensor
  • Iniciar sensor con el código “real” (impreso en el protector adhesivo)

Check in xDrip+ logs if xDrip+ starts counting “Duration: 1 minute” (and so on). Only in the xDrip+ logs you can detect at an early stage whether xdrip+ has stopped a sensor. Latest status is not always shown correctly on bottom of startscreen.

xDrip+ & Freestyle Libre

Libre configuración específica

  • Abrir la Configuración Bluetooth -> Menú Hamburguesa (parte superior izquierda de la pantalla) -> Configuración -> desplácese hacia abajo -> Ajustes menos comunes -> Ajustes de Bluetooth

    xDrip+ Libre Configuración 1 de Bluetooth

  • Habilite los siguientes parámetros

    • Encender Bluetooth
    • Usar escaneo
    • Siempre descubrir servicios
  • Todas las demás opciones deben ser desactivados

    xDrip+ Libre Configuración 2 de Bluetooth

Conectar el transmisor de Libre & iniciar sensor

xDrip+ Start Libre Transmitter & Sensor 1

xDrip+ Start Libre Transmitter & Sensor 2

xDrip+ Start Libre Transmitter & Sensor 3