Equil
These instructions are for configuring the Equil insulin pump.
Pump capabilities with AAPS
Tubeless patch pump that communicates with AAPS over your phone’s native Bluetooth, without an additional communication device.
AAPS treats the Equil as a patch pump with a 200 U reservoir: the main screen shows the patch pump status lights, and Prime/Fill and Pump battery change are not offered. Filling, priming and changing the battery are done in the pump wizards.
While the pump is suspended or stopped, AAPS records a 0 U/h temporary basal, so no basal insulin is counted in IOB.
DST and timezone changes must be handled manually.
Hardware and software requirements
Compatible Equil hardware
Currently Equil 5.3 and 5.4 are supported.
AAPS built and installed using the Build APK instructions
Setup
Select Equil pump
Open the menu (☰) in the top-left corner, choose Configuration, and under Pump select Equil.
Once it is selected, the Equil card shows two buttons, Settings and Open plugin:

Open plugin (or Manage → Pump) opens the Equil pump screen. Before a pump is paired it looks like this:

Settings
Open the Equil settings with Settings on the Equil card, or with the gear icon in the top-right corner of the pump screen.

Low battery alarm: warns you when the pump battery is low.
Low drug storage alarm: warns you when the insulin in the reservoir is low.
Alert tone: how the pump alerts you: Mute, Vibrate only, Tone only or Vibrate and tone.
Bolus speed: how fast a bolus is given: 1.5 U/min (standard), 3 U/min, 6 U/min or 12 U/min. A faster bolus may give more occlusion alarms.

Max basal and max bolus on the patch
The patch itself refuses any basal rate above a maximum that AAPS programs into it. AAPS takes the higher of your max basal setting and the highest hourly basal rate in your Profile, so the patch always accepts your profile and the temp basals the loop asks for. AAPS also sends your maximum allowed bolus to the patch.
When you change your max basal or max bolus, AAPS sends the new values to the patch automatically. They are also sent with every profile switch. There is nothing to set on the patch itself.
Pair the pump
On the pump screen, press Pair. A wizard guides you through each step. The row of dots at the top shows how far along you are.
If Android asks whether AAPS may find and connect to nearby devices, press Allow. AAPS needs this to find the pump.

The wizard has these steps. Press Next to move on, or Cancel to leave the wizard.
Profile required: only shown if you have never activated a Profile in AAPS. Select the profile to use and press Activate profile.
Assemble pump: put together the pump, the filled reservoir and the charged battery.

Find Pump: AAPS scans for nearby Equil pumps. Select your pump from the list.

Pair device: you can enter a pairing password in Set pair password. It must be 4 characters long, chosen from
ABCDEF0123456789. Press Pair.
Select Insulin: only shown if you have more than one insulin set up. Select the insulin in the reservoir.
Prime / Fill: the pump must not be on its base plate. Press Prime/Fill and prime the reservoir until there is a drop of insulin on the needle tip. Press Next.

Site Rotation: only shown if Manage pump site rotation is turned on in the site rotation settings. Tap where you will attach the pump, or press Skip.
Attach pump: attach the pump to its base plate.

Prime cannula: once the pump is on the base plate, press Purge air to remove air from the cannula. Press Next.

Confirm: put the pump on your body and press Finish.

Warning
If you set your own pairing password (recommended for your safety), store it somewhere safe. The password is saved in the pump. The pump asks for it at every new pairing until you unpair it properly in AAPS. Until then, the pump also cannot be used with the original handheld controller (PDA).
Pump screen after pairing
Once the pump is paired, the pump screen shows these buttons:
Refresh: reads the state and the history of the pump now.
Suspend / Resume delivery: stop or restart insulin delivery.
Change reservoir: guides you through detaching the pump, fitting a new reservoir (and, if needed, a new battery), priming and attaching it again. The first step asks Will you also put on a new infusion set (base plate and cannula)?
Yes, new infusion set: you choose the site, the cannula is primed, and AAPS records an insulin change and a cannula change.
No, only reservoir or battery: the cannula is not primed and AAPS records only an insulin change.
After a reservoir change the battery level shows - until AAPS reads it again from the pump.

History: shows the events recorded by the pump.
Unpair device: guides you through detaching the pump and unpairing it from AAPS. This stops insulin delivery, moves the pump plunger back to the bottom, and deletes the pairing.
Alarms
The low insulin alarm sounds once at each level (10 U, 5 U and 2 U) and when the reservoir is empty. The low battery alarm sounds once at 10% and once at 5%. Each level alarms only once until the next reservoir change.
Pump alarms and stops are also recorded as announcements in the treatments, at the time the pump logged them, so you can see later when and why the pump stopped.
If the reservoir runs empty during a bolus, AAPS records only the insulin given until then and tells you both amounts.
Where to get help
Development of the Equil driver is done by the community on a volunteer basis. Before requesting help, please:
Read the relevant section of this documentation to confirm how the feature is meant to work.
Ask on the #AAPS channel on Discord, or in one of the other community channels.
Report a bug by searching the existing issues; if yours is not listed, open a new issue and attach your log files.
When asking for help, include your phone make and model, Android version, AAPS version, and a plain-English description of the problem (what changed, when it last worked).