xDrip settings

If not already set up, download and install the latest xDrip release from the xDrip releases page.

Disable battery optimization and allow background activity for the xDrip app.

Basic settings for all CGM & FGM systems

Disable Nightscout upload

You shouldn’t let any other app upload data (blood glucose and treatments) to Nightscout.

→ Hamburger Menu (1) → Settings (2) → Cloud Upload (3) → Nightscout Sync (REST-API) (4) → Switch OFF Enabled (5)

xDrip Basic Settings 1

Disable automatic calibration and treatments

Make sure Automatic Calibration (7) is deactivated. If the checkbox for Automatic Calibration is checked, activate Download treatments (6) once, then remove the checkbox for Automatic Calibration and deactivate Download treatments again.

xDrip Basic Settings 2

Tap Extra Options(8)

Safety warning

You must deactivate “Upload treatments” from xDrip, otherwise treatments can be doubled in AAPS leading to false COB and IOB.

Deactivate Upload treatments(9) and make sure you will NOT use Back-fill data (11).

Option Alert on failures should also be deactivated (10). Otherwise you will get an alarm every 5 minutes in case Wi-Fi/mobile network issues or if the server is not available.

xDrip Basic Settings 3

Inter-app Settings (Broadcast)

If you use AAPS, you must activate broadcasting in xDrip’s Inter-App settings so the data is forwarded to AAPS.

→ Hamburger Menu (1) → Settings (2) → Inter-app settings (3) → Broadcast locally ON (4)

In order for the values to be identical in AAPS with respect to xDrip, you should activate Send the displayed glucose value (5).

Enable Compatible Broadcast (6).

xDrip Basic Settings 4

If you have also activated Accept treatments in xDrip and Enable broadcasts to xDrip in AAPS xDrip plugin, then xDrip will receive insulin, carbs and basal rate information from AAPS.

If you enable Accept Calibrations, xDrip will use the calibrations from AAPS. Be careful when you use this feature with Dexcom sensors: read this first.

Remember to disable Import Sounds to avoid xDrip making a ringtone every time AAPS sends a basal/profile change.

xDrip Basic Settings 5

Identify receiver

  • If you discover problems with local broadcast (AAPS not receiving BG values from xDrip) go to → Hamburger Menu (1) Settings (2) → Inter-app settings (3) → Identify receiver (7) and enter info.nightscout.androidaps for AAPS build (if you are using PumpControl build, please enter info.nightscout.aapspumpcontrol instead!!).

  • Pay attention: Auto-correction sometimes tend to change i to capital letter. You must use only lowercase letters when typing info.nightscout.androidaps (or info.nightscout.aapspumpcontrol for PumpControl). Capital I would prevent the App from receiving BG values from xDrip.

    xDrip Basic Inter-app Settings Identify receiver

Use AAPS to calibrate in xDrip

  • If you want to be able to use AAPS to calibrate then in xDrip go to Settings → Interapp Compatibility → Accept Calibrations and select ON.

  • You may also want to review the options in Settings → Less Common Settings → Advanced Calibration Settings.

Dexcom sensors

Everything related to the Dexcom sensor and transmitter themselves (starting, stopping, replacing, calibrating, troubleshooting) is documented and kept up to date in the xDrip Dexcom documentation. Follow those instructions for the sensor; the settings connecting xDrip to AAPS (described above on this page) are the only AAPS-specific part.

Dexcom G6 and ONE

  • Set up xDrip following the recommended settings.

  • The Dexcom G6 and ONE transmitters can simultaneously be connected to the Dexcom receiver (or alternatively the t:slim pump) and one app on your phone.

  • When using xDrip as receiver uninstall the Dexcom app first. You cannot connect xDrip and the Dexcom app with the transmitter at the same time!

  • If you need Clarity and want to profit from xDrip features, use the Build Your Own Dexcom App with local broadcast to xDrip, or use xDrip as a Companion app receiving notifications from the official Dexcom app (with possible delays in BG readings).

Sensor restarts

  • xDrip always uses the transmitter’s native algorithm with current transmitters; there is no reason to disable it.

  • Preemptive restarts only function on old transmitters (firmware 1.6.5.25 or earlier), where they are needed to prevent the sensor stopping after 10 days. Recent (Firefly) transmitters ignore them.

  • Restarting a sensor on a Firefly transmitter requires removing the transmitter and readings can be significantly in error after the restart, which is dangerous when looping: read Restart a sensor in full and calibrate carefully.

Start, stop and replace sensors

Follow the instructions:

Replace a transmitter

  • Turn the original Dexcom receiver off (if used) and do not turn it back on before xDrip shows readings.

  • Forget the device in xDrip (System Status → Forget Device) AND in the smartphone’s Bluetooth settings (shown as DexcomXX where XX are the last two digits of the transmitter serial number).

  • Keep the old transmitter out of Bluetooth range to prevent reconnection.

  • Then follow Start a G6 or ONE transmitter.

Dexcom G7, ONE+ and Stelo

Troubleshooting Dexcom and xDrip

Libre 1

  • Setup your NFC to Bluetooth bridge in xDrip

    → Hamburger Menu (1) → Settings (2) → Less common settings (3) → Bluetooth Settings (4)

  • In Bluetooth Settings set the checkboxes exactly as in the screenshots below (5)

    • Disable watchdogs as they will reset the phone Bluetooth and interrupt your pump connection.

    xDrip Libre Bluetooth Settings 1

  • You can try to enable the following settings (7)

    • Use scanning

    • Trust Auto-Connect

    • Use Background Scans

  • If you easily lose connection to the bridge or have difficulties recovering connection, DISABLE THEM (8).

    xDrip Libre Bluetooth Settings 2

  • Leave all other options disabled unless you know why you want to enable them.

    xDrip Libre Bluetooth Settings 3

Libre smart reader battery level

  • Battery level of bridges such as MiaoMiao and Bubble can be displayed in AAPS (not Blucon).

  • Details can be found on screenshots page.

Connect Libre Transmitter & start sensor

  • If your sensor requires it (Libre 2 EU and Libre 1 US) install the latest out of process algorithm.

  • Your sensor must be already started using the vendor app or the reader (xDrip cannot start or stop Libre sensors).

  • Set the data source to Libre Bluetooth.

    → Hamburger Menu (1) → Settings (2) → Select Libre Bluetooth in Hardware Data source (3)

    xDrip Start Libre Transmitter & Sensor 1

  • Scan Bluetooth and connect the bridge.

    → Hamburger Menu (1) → Scan Bluetooth (2) → Scan (3)

    • If xDrip can’t find the bridge, make sure it’s not connected to the vendor app. Put it in charge and reset it.

    xDrip Start Libre Transmitter & Sensor 2

  • Start the sensor in xDrip.

    Safety warning

    Do not use sensor data before the one hour warm-up is over: the values can be extremely high and cause wrong decisions in AAPS.

    → Hamburger Menu (1) → Start sensor (2) → Start sensor (3) → Set the exact time you started it with the reader or the vendor app. If you didn’t start it today, answer “Not Today” (4).

xDrip Start Libre Transmitter & Sensor 3

Libre 2 patched app

  • Set the data source to Libre patched app.

    → Hamburger Menu (1) → Settings (2) → Select Libre (patched App) in Hardware Data source (3)

    xDrip Libre Patched app 1

  • You can add BgReading:d,xdrip libre_receiver:v under Less Common Settings->Extra Logging Settings->Extra tags for logging. This will log additional error messages for trouble shooting.

xDrip LibreLink logging