xDrip settings
If not already set up, download and install the latest xDrip release from the xDrip releases page.
Disable battery optimization and allow background activity for the xDrip app.
Basic settings for all CGM & FGM systems
Disable Nightscout upload
You shouldn’t let any other app upload data (blood glucose and treatments) to Nightscout.
→ Hamburger Menu (1) → Settings (2) → Cloud Upload (3) → Nightscout Sync (REST-API) (4) → Switch OFF Enabled (5)

Disable automatic calibration and treatments
Make sure Automatic Calibration (7) is deactivated.
If the checkbox for Automatic Calibration is checked, activate Download treatments (6) once, then remove the checkbox for Automatic Calibration and deactivate Download treatments again.

Tap Extra Options(8)
Safety warning
You must deactivate “Upload treatments” from xDrip, otherwise treatments can be doubled in AAPS leading to false COB and IOB.
Deactivate Upload treatments(9) and make sure you will NOT use Back-fill data (11).
Option Alert on failures should also be deactivated (10). Otherwise you will get an alarm every 5 minutes in case Wi-Fi/mobile network issues or if the server is not available.

Inter-app Settings (Broadcast)
If you use AAPS, you must activate broadcasting in xDrip’s Inter-App settings so the data is forwarded to AAPS.
→ Hamburger Menu (1) → Settings (2) → Inter-app settings (3) → Broadcast locally ON (4)
In order for the values to be identical in AAPS with respect to xDrip, you should activate Send the displayed glucose value (5).
Enable Compatible Broadcast (6).

If you have also activated Accept treatments in xDrip and Enable broadcasts to xDrip in AAPS xDrip plugin, then xDrip will receive insulin, carbs and basal rate information from AAPS.
If you enable Accept Calibrations, xDrip will use the calibrations from AAPS. Be careful when you use this feature with Dexcom sensors: read this first.
Remember to disable Import Sounds to avoid xDrip making a ringtone every time AAPS sends a basal/profile change.

Identify receiver
If you discover problems with local broadcast (AAPS not receiving BG values from xDrip) go to → Hamburger Menu (1) Settings (2) → Inter-app settings (3) → Identify receiver (7) and enter
info.nightscout.androidapsfor AAPS build (if you are using PumpControl build, please enterinfo.nightscout.aapspumpcontrolinstead!!).Pay attention: Auto-correction sometimes tend to change i to capital letter. You must use only lowercase letters when typing
info.nightscout.androidaps(orinfo.nightscout.aapspumpcontrolfor PumpControl). Capital I would prevent the App from receiving BG values from xDrip.
Use AAPS to calibrate in xDrip
If you want to be able to use AAPS to calibrate then in xDrip go to Settings → Interapp Compatibility → Accept Calibrations and select ON.
You may also want to review the options in Settings → Less Common Settings → Advanced Calibration Settings.
Dexcom sensors
Everything related to the Dexcom sensor and transmitter themselves (starting, stopping, replacing, calibrating, troubleshooting) is documented and kept up to date in the xDrip Dexcom documentation. Follow those instructions for the sensor; the settings connecting xDrip to AAPS (described above on this page) are the only AAPS-specific part.
Dexcom G6 and ONE
Set up xDrip following the recommended settings.
The Dexcom G6 and ONE transmitters can simultaneously be connected to the Dexcom receiver (or alternatively the t:slim pump) and one app on your phone.
When using xDrip as receiver uninstall the Dexcom app first. You cannot connect xDrip and the Dexcom app with the transmitter at the same time!
If you need Clarity and want to profit from xDrip features, use the Build Your Own Dexcom App with local broadcast to xDrip, or use xDrip as a Companion app receiving notifications from the official Dexcom app (with possible delays in BG readings).
Sensor restarts
xDrip always uses the transmitter’s native algorithm with current transmitters; there is no reason to disable it.
Preemptive restarts only function on old transmitters (firmware 1.6.5.25 or earlier), where they are needed to prevent the sensor stopping after 10 days. Recent (Firefly) transmitters ignore them.
Restarting a sensor on a Firefly transmitter requires removing the transmitter and readings can be significantly in error after the restart, which is dangerous when looping: read Restart a sensor in full and calibrate carefully.
Start, stop and replace sensors
Follow the instructions:
Transmitter lifetime can only be extended (hard reset) on rebatteried or modified transmitters, not on Firefly.
Replace a transmitter
Turn the original Dexcom receiver off (if used) and do not turn it back on before xDrip shows readings.
Forget the device in xDrip (System Status → Forget Device) AND in the smartphone’s Bluetooth settings (shown as DexcomXX where XX are the last two digits of the transmitter serial number).
Keep the old transmitter out of Bluetooth range to prevent reconnection.
Then follow Start a G6 or ONE transmitter.
Dexcom G7, ONE+ and Stelo
Follow the G7, ONE+ and Stelo instructions (a recent xDrip release is required — see the linked page).
See also the Dexcom G7, ONE+ and Stelo page for the available setups with AAPS.
Troubleshooting Dexcom and xDrip
Full list: see Troubleshooting on the xDrip Dexcom page.
Libre 1
Setup your NFC to Bluetooth bridge in xDrip
→ Hamburger Menu (1) → Settings (2) → Less common settings (3) → Bluetooth Settings (4)
In Bluetooth Settings set the checkboxes exactly as in the screenshots below (5)
Disable watchdogs as they will reset the phone Bluetooth and interrupt your pump connection.

You can try to enable the following settings (7)
Use scanning
Trust Auto-Connect
Use Background Scans
If you easily lose connection to the bridge or have difficulties recovering connection, DISABLE THEM (8).

Leave all other options disabled unless you know why you want to enable them.

Libre smart reader battery level
Battery level of bridges such as MiaoMiao and Bubble can be displayed in AAPS (not Blucon).
Details can be found on screenshots page.
Connect Libre Transmitter & start sensor
If your sensor requires it (Libre 2 EU and Libre 1 US) install the latest out of process algorithm.
Your sensor must be already started using the vendor app or the reader (xDrip cannot start or stop Libre sensors).
Set the data source to Libre Bluetooth.
→ Hamburger Menu (1) → Settings (2) → Select Libre Bluetooth in Hardware Data source (3)

Scan Bluetooth and connect the bridge.
→ Hamburger Menu (1) → Scan Bluetooth (2) → Scan (3)
If xDrip can’t find the bridge, make sure it’s not connected to the vendor app. Put it in charge and reset it.

Start the sensor in xDrip.
Safety warning
Do not use sensor data before the one hour warm-up is over: the values can be extremely high and cause wrong decisions in AAPS.
→ Hamburger Menu (1) → Start sensor (2) → Start sensor (3) → Set the exact time you started it with the reader or the vendor app. If you didn’t start it today, answer “Not Today” (4).

Libre 2 patched app
Set the data source to Libre patched app.
→ Hamburger Menu (1) → Settings (2) → Select Libre (patched App) in Hardware Data source (3)

You can add
BgReading:d,xdrip libre_receiver:vunder Less Common Settings->Extra Logging Settings->Extra tags for logging. This will log additional error messages for trouble shooting.