EOPatch2

The patch needs rapid-acting U-100 insulin, such as NovoRapid or Humalog. Use the rapid-acting insulin your doctor has prescribed for you.

The smallest dose the patch can deliver is 0.05 U. Set every basal rate in your Profile to at least 0.05 U/hr, in steps of 0.05 U/hr. Otherwise AAPS rounds each rate to the nearest 0.05 U/hr (never down to 0) and uses the rates the patch really gives, so the total insulin per day differs slightly from your Profile. In the same way, boluses must be at least 0.05 U.

Pump capabilities with AAPS

  • Communicates with AAPS over your phone’s native Bluetooth, without an additional communication device.

  • DST and timezone changes must be handled manually.

Pump setup

  1. Open the menu (☰) in the top-left corner of the AAPS main screen and select Configuration.

  2. In the Pump section, select EOPatch2.

  3. Press the Back key to return to the main screen.

The menu with the Configuration entry Configuration > Pump with the EOPatch2 plugin

Once it is selected, the EOPatch2 card shows two buttons, Settings and Open plugin:

The EOPatch2 plugin selected in Configuration > Pump

Settings

Open the EOPatch2 settings in one of two ways:

  • press Settings on the EOPatch2 card in Configuration, or

  • open the pump screen (Manage → Pump) and press the gear icon in the top-right corner.

There are three settings:

EOPatch2 settings

Low reservoir Alerts

A warning appears when the insulin left in the patch drops to this amount or less. You can set it from 10 to 50 U, in steps of 5 U. The default is 10 U.

Patch expiration Reminder

Reminds you how much time is left before the current patch expires. You can set it from 1 to 24 hours, in steps of 1 hour. The default is 4 hours.

Patch buzzer Reminders

When this is on, the patch beeps when a bolus, extended bolus or temporary basal starts and when it ends. It has no effect on your normal basal. The default is off.

Activating a new patch

Open the activation wizard

Open the pump screen with Manage → Pump (or press Open plugin on the EOPatch2 card in Configuration). When no patch is active, the screen shows Not activated and an Activate Patch button at the bottom.

Press Activate Patch to start the wizard.

The EOPatch2 pump screen with the Activate Patch button

The wizard guides you through each step. The row of dots at the top shows how far along you are.

Note

If you have never activated a Profile in AAPS, the wizard first shows a Profile required page. Select the profile to use and press Activate profile.

Filling Insulin

Insert the syringe needle into the insulin fill port on the patch. Slowly push the plunger to fill the patch. Once there is more than 80 U in the patch, it beeps once and starts up (boots).

When you hear the beep, press Start pairing.

The Filling Insulin step with the Start pairing button

Warning

  • Do not turn the needle action lever until the wizard tells you to. Turning it too early can cause serious problems during filling or the safety check.

  • The patch holds 80 to 200 U. If you put in less than 80 U, the patch does not work.

  • Take the insulin out of the refrigerator 15 to 30 minutes before filling, so it reaches room temperature. The insulin must be at least 10°C.

Patch pairing

AAPS now tries to pair with the patch automatically. During pairing, two Android Bluetooth pairing requests appear. Press OK on the first one. Press OK again on the second one, which shows a passkey.

Warning

  • Keep the patch and your phone within 30 cm of each other while pairing.

  • After the patch has booted, it beeps every 3 minutes until pairing is complete.

  • You must finish activating the patch within 60 minutes of it booting. If you cannot, discard the patch.

Patch pairing step First Bluetooth pairing request Second Bluetooth pairing request with the passkey

Older screenshot

The two Bluetooth pairing screenshots are from an earlier AAPS version: the AAPS screen behind the Android pairing request looks different in AAPS 4.

Select insulin

Select the insulin you filled the patch with and press Next. AAPS applies a profile switch with this insulin after activation.

Prepare for attaching the Patch

Remove the adhesive tape from the patch, then press Next.

If a needle sticks out, or if the patch is wet, dirty or its adhesive tape is folded, press Discard instead and use a new patch.

Prepare for attaching the Patch step

Site location

This step only appears if Manage pump site rotation is turned on in the site rotation settings. Tap the place on the body diagram where you will attach the patch and press Next, or press Skip.

Attaching the Patch

Insulin should go into a spot with fatty tissue under the skin, but few nerves or blood vessels. The abdomen, arm or thigh are good choices. Clean and dry the site, then attach the patch to your skin.

Check the infusion site, then press Start safety check.

Warning

  • Press down the edges of the patch tape evenly, so the whole patch sticks firmly to your skin.

  • If the patch does not stick completely, air can get between the patch and your skin. This weakens the adhesive and the water resistance of the patch.

Attaching the Patch step

Safety check

The safety check takes about 30 seconds. When it is finished, the patch beeps once. If it fails, press Retry.

Warning

For safe use, do not turn the needle action lever until the safety check is complete.

Attaching the Patch step with the Start safety check button Safety check in progress

Older screenshot

The screenshots from Prepare for attaching the Patch to Safety check are from an earlier AAPS version and may look different in AAPS 4.

Inserting the needle

Hold the patch firmly. Turn the needle action lever upward by more than 100° to insert the needle. The patch beeps once when the needle is inserted correctly. Then keep turning the lever all the way to remove it. Press Next.

Caution

If you go to the next step without hearing the beep, a Needle insertion Error appears. Check for the beep and press Retry. If the error remains, press Discard to deactivate the patch.

Patch activation completed

The last page confirms that the patch is active and that it will remind you when it nears its expiration time. Press Finish to go back to the pump screen.

Discarding the patch

Replace the patch when it is low on insulin, when it expires, or if it is faulty. We recommend using each patch for no more than 84 hours after it boots.

  1. Open the pump screen (Manage → Pump). While a patch is active, press Discard Patch.

  2. The Discard Patch page shows the time left and the insulin left in the patch. Press Discard Patch.

  3. A dialog asks you to confirm. Press Discard Patch again.

  4. When the patch has been deactivated, the wizard shows Discarding the patch is completed. Remove the patch from your body and press Confirm.

After that, the wizard continues straight on to activating a new patch (see above). Press the back arrow if you want to stop here.

While no patch is active, AAPS shows the pump as suspended and records a 0 U/hr temporary basal, so no basal insulin is counted in IOB until the next patch is activated.

The EOPatch2 pump screen with the DISCARD/CHANGE PATCH button Discard Patch page Discard Patch confirmation dialog Discarding the patch is completed

Older screenshot

The screenshots above are from an earlier AAPS version. In AAPS 4 you open the pump screen from Manage → Pump, and the button on it is labelled Discard Patch.

If the patch does not respond

If AAPS cannot reach the patch, it may show a page about turning off the patch’s alarm by hand: remove the patch from your body, peel off the adhesive tape, and use a clip to press firmly into the hole next to the insulin fill port.

The Discard Patch page also has a Force Reset button. It only clears the patch from AAPS, without sending any command to the patch. The patch keeps delivering insulin until its battery runs out or you turn it off by hand. Use it only if a normal discard fails, and remove the patch from your body.

Suspending and resuming insulin delivery

Suspending insulin delivery also cancels any extended bolus and temporary basal. When you resume, these are not restarted. While insulin delivery is suspended, the patch beeps every 15 minutes.

Suspending insulin delivery

  1. Open the pump screen (Manage → Pump) and press Suspend pump.

  2. A dialog explains what will be suspended. Press Confirm.

  3. In the Basal Suspending Time dialog, choose how long to suspend: 30 min, 1 hr, 1 hr 30 min or 2 hr. Press Confirm.

Insulin delivery is suspended for the time you chose.

The EOPatch2 pump screen with the SUSPEND button Suspend confirmation dialog Basal Suspending Time dialog

Older screenshot

The screenshots above are from an earlier AAPS version. In AAPS 4 you open the pump screen from Manage → Pump, and the button on it is labelled Suspend pump.

Resuming insulin delivery

  1. Open the pump screen (Manage → Pump) and press Resume pump.

  2. In the Resume insulin delivery dialog, press Confirm.

The EOPatch2 pump screen with the RESUME button Resume insulin delivery dialog

Older screenshot

The screenshots above are from an earlier AAPS version. In AAPS 4 you open the pump screen from Manage → Pump, and the button on it is labelled Resume pump.

Warning

When the suspend time ends, AAPS shows an End of insulin suspend warning. Insulin delivery does not restart until you confirm it. If you do not, you could develop hyperglycemia.

Alarms and warnings

Alarms

Alarms are for urgent, top-priority situations and need your action straight away. AAPS sounds an alarm that keeps going until you acknowledge it, and shows the message as a notification on the Overview screen. An alarm means there is a problem with the patch in use. In many cases the patch has to be discarded and replaced with a new one.

When you press Confirm on the notification, AAPS handles the alarm. For most alarms, this deactivates the patch. For an Inappropriate temperature alarm, the button is Retry.

EOPatch2 alarm EOPatch2 alarm after the patch was deactivated

Older screenshot

The alarm screenshots are from an earlier AAPS version and may look different in AAPS 4.

The different alarms are explained below.

Alarm

Explanation

Empty reservoir

The patch has run out of insulin.

Patch expired

The patch’s usage time is over, and it cannot deliver any more insulin.

Occlusion

The patch’s insulin path seems to be blocked.

Power on self-test failure

The patch found an unexpected error during its self-test after booting.

Inappropriate temperature

The patch is outside its normal operating temperature while you activate or use it. Move the patch to a place with a suitable temperature (4.4 to 37°C).

Needle insertion Error

The needle was not inserted correctly during activation. Check that the needle insertion edge of the patch and the needle action lever are in a straight line.

Patch battery Error

The patch’s internal battery is about to run out and the patch will power off.

Patch activation Error

The patch was not fully activated within 60 minutes of booting.

Patch Error

The patch found an unexpected error while being activated or used.

Warnings

Warnings are for medium- or low-priority situations. A warning appears as a notification on the Overview screen.

EOPatch2 warning shown on the Overview screen

Older screenshot

This screenshot is from an earlier AAPS version: in AAPS 4 the warning appears on the main screen, which looks different.

The different warnings are explained below.

Warning

Explanation

End of insulin suspend

The suspend time you chose is over. Press Resume pump on the notification to restart insulin delivery.

Low reservoir

The insulin left in the patch is below the amount set in Low reservoir Alerts.

Patch operating life expired

The patch’s usage period is over.

Patch will expire soon

The patch must be discarded in 1 hour.

Incomplete Patch activation

Patch activation was interrupted for more than 3 minutes after pairing. Open the pump screen and finish activating the patch.

Patch battery low

The patch’s battery is low.

Where to get help

Development of the EOPatch2 driver is done by the community on a volunteer basis. Before requesting help, please:

  1. Read the relevant section of this documentation to confirm how the feature is meant to work.

  2. Ask on the #AAPS channel on Discord, or in one of the other community channels.

  3. Report a bug by searching the existing issues; if yours is not listed, open a new issue and attach your log files.

When asking for help, include your phone make and model, Android version, AAPS version, and a plain-English description of the problem (what changed, when it last worked).