Konfigurations-Generator
Depending on your settings you can open Config Builder through a tab at the top of AAPS‘ screen or through the hamburger menu.
The Config Builder is the tab where you turn the modular features on and off. In the picture below, the boxes on the left-hand side (A) allow you to select which modules you want activated, the boxes on the right-hand side (C) allow you to view these as a tab (E) in AAPS. In case the right box is not activated, you can reach the function by using the hamburger menu (D) on the top left of the screen. See Tab or hamburger menu below.
When there are additional settings available within the module, you can click on the cog wheel (B) which will take you to the specific settings within preferences.
Profile
This module can not be disabled as it is a core part of AAPS.
See Your AAPS Profile for a basic understanding of what goes inside your Profile.
See AAPS Screens > Profile for more information about managing your Profiles.
Insulin
Select the type of insulin you are using.
More information to understand the Insulin Profile as shown in AAPS here.
Unterschiede der Insulintypen
Die Oref Optionen ‚Rapid-Acting Oref‘, Ultra-Rapid Oref‘, ‚Lyumjev‘ und ‚Free-Peak Oref‘ haben einen exponentiellen Verlauf.
Bei den Profilen ‚Rapid-Acting‘, ‚Ultra-Rapid‘ und ‚Lyumjev‘ kannst du nur die Insulinwirkdauer (DIA) anpassen. Der Zeitpunkt der maximalen Insulinwirkung ist fest vorgegeben.
Das Profil Free-Peak erlaubt, nicht nur die Insulinwirkdauer (DIA), sondern auch den Zeitpunkt der maximalen Insulinwirkung individuell festzulegen. Es sollte nur von erfahrenen Anwendern genutzt werden, die die Auswirkungen dieser Einstellungen kennen.
The insulin curve graph helps you to understand the different curves.
Rapid-Acting Oref
empfohlen für Humalog, Novolog und Novorapid
DIA = mindestens 5 Stunden
Wirkmaximum = 75 Minuten nach Insulingabe (fest eingestellt, nicht anpassbar)
Ultra-Rapid Oref
empfohlen für FIASP
DIA = mindestens 5 Stunden
Wirkmaximum = 55 Minuten nach Insulingabe (fest eingestellt, nicht anpassbar)
Lyumjev
Spezielles Insulinprofil für Lyumjev
DIA = mindestens 5 Stunden
Wirkmaximum = 45 Minuten nach Insulingabe (fest eingestellt, nicht anpassbar)
Free Peak Oref
Erlaubt das Wirkmaximum der Insulinaktivität individuell zu definieren. Klicke dazu auf das Zahnrad um erweiterte Einstellungen einzugeben.
DIA wird automatisch auf 5 Stunden gesetzt, sofern von dir im Profil nichts anderes definiert wird.
Dieses Profil wird dann empfohlen, wenn dein Insulintyp von den anderen Profilen nicht abgedeckt werden kann oder wenn eine Mischung verschiedener Insuline verwendet wird.
BZ-Quelle
Select the blood glucose source you are using. See BG Source page for more setup information.
NSClient BG - only if you know what you are doing, see BG Source.
Glimp - only version 4.15.57 and newer are supported
Tomato App for MiaoMiao device
Glunovo App für Glunovo CGM System
Zufalls-BZ: Generiert zufällige Blutzuckerdaten (nur im Demo-Modus)
Smoothing
Pump
Wähle die von dir genutzte Pumpe. See Compatible pumps page for more setup information.
DanaR Korean (koreanische Version der DanaR)
Dana Rv2 (DanaR mit inoffiziellem Firmware Upgrade)
Accu Chek Combo
Driver using Ruffy (requires ruffy installation)
Omnipod for Omnipod Eros
Dash for Omnipod DASH
Virtual pump: open loop - AAPS suggestions only
as you make you first steps with AAPS, during the first objectives
for pump which doesn’t have any driver yet
Empfindlichkeitserkennung
Hier kannst Du auswählen, mit welchem Algorithmus AAPS die Insulinempfindlichkeit berechnen soll. For more details of different designs please read on here. Bei der Empfindlichkeitserkennung werden historische Daten „on the go“ analysiert und Anpassungen vorgenommen, falls der Algorithmus feststellt, dass du sensibler oder weniger empfindlich auf das Insulin reagierst als üblich. Mehr Details zum Sensitivitäts Algorithmus findest du in den OpenAPS Docs.
You can view your sensitivity on the homescreen in an additional graph. Die berechnete Insulinempfindlichkeit kannst du verfolgen, indem du auf der Startseite im Auswahlmenü der angezeigten Kurven “Sensitivität” auswählst. Die weiße Linie zeigt dir das graphisch an. Note, you need to be in Objective 8 in order to let Sensitivity Detection/Autosens automatically adjust the amount of insulin delivered. So lange Du dieses Ziel (objective) noch nicht erreicht hast, wird der Autosens-Prozentsatz bzw. die Autosens-Kurve nur zu Deiner Information angezeigt. AAPS nimmt keine Änderungen vor.
Resorptions-Einstellungen
If you use Oref1 with SMB you must change min_5m_carbimpact to 8. The value is only used during gaps in CGM readings or when physical activity „uses up“ all the blood glucose rise that would otherwise cause AAPS to decay COB. At times when carb absorption can’t be dynamically worked out based on your blood’s reactions it inserts a default decay to your carbs. Im Prinzip ist es eine Notlauffunktion.
APS
Wähle den gewünschten APS-Algorithmus für Therapie-Anpassungen. Die Details zum ausgewählten Algorithmus findest du im Reiter OpenAPS (OPAS).
OpenAPS AMA
Advanced Meal Assist: older algorithm not recommended anymore.
In simple terms, the benefits are after you give yourself a meal bolus, the system can high-temp more quickly IF you enter carbs reliably.
-
Super Micro Bolus: most recent algorithm recommended for all users.
In contrast to AMA, SMB does not use temporary basal rates to control glucose levels, but mainly small Super Micro Boluses.
Note : It is recommended to use this algorithm from the beginning, even though you will not actually get SMBs delivered until Objective 9.
If switching from AMA to SMB algorithm, min_5m_carbimpact must be changed manually to 8 (default value for SMB) in Preferences > Sensitivity detection > Sensitivity Oref1 settings.
Loop
This module should not be disabled as it is a core part of AAPS.
Constraints
Objectives
AAPS has a learning program (a series of objectives) that you have to fulfill step by step. Dies soll dich sicher durch die Einrichtung eines Closed Loop Systems führen. Das garantiert, dass du alles korrekt eingestellt hast und auch verstehst, was das System genau macht. Nur so kannst du dem System vertrauen.
See Objectives page for more information.
Synchronisierung
In this section, you can choose if/where you want AAPS to send your data to.
NSClient or NSClientV3
Can be used as a reporting server and/or for remote monitoring, remote control.
See Synchronization with the reporting server to help you choose between NSClient (v1) and NSClientV3.
Tidepool
Can be used as a reporting server.
See Tidepool.
xDrip
Used to send data such as treatments to xDrip+.
Open Humans
See Open Humans.
Wear
Monitor and control AAPS using your Android WearOS watch (see page Watchfaces).
Samsung Tizen
Broadcast data to Samsung’s G-Watch Wear App (Tizen OS).
Garmin
Connection to Garmin device (Fenix, Edge…)
Behandlungen
Der Reiter Behandlungen (BEH) zeigt Dir die zu Nightscout hochgeladenen Behandlungen. Should you wish to edit or delete an entry (e.g. you ate less carbs than you expected) then select ‚Remove‘ and enter the new value (change the time if necessary) through the carbs button on the home screen.
General
Übersicht
This is the main screen of AAPS and can not be disabled.
Zeige Feld für Notizen in den Behandlungsdialogen
Hier kannst Du das Notizfeld für die Behandlungsdialoge (Bolus-Rechner, Insulin, Kohlenhydrate) ein- oder ausschalten.
Statusanzeige
Choose if you want to have status lights on overview for cannula age, insulin age, sensor age, battery age, reservoir level or battery level. Bei Überschreiten der Warnschwelle werden die Werte gelb angezeigt. Ab der kritischen Warnschwelle werden die Werte rot angezeigt.
Erweiterte Einstellungen
Abgabe nur eines Teils der vom Bolus-Rechner ermittelten Insulinmenge: Viele Anwender geben bei der Nutzung von SMB nicht mehr 100% der vom Bolus-Rechner ermittelten Insulinmenge ab. Stattdessen geben Sie nur einen Teil (z.B. 75%) ab und lassen SMB und UAM („nicht angemeldete Mahlzeiten“) den Rest erledigen. In this setting, you can choose a default value for the percentage the bolus wizard should calculate with. Wenn Du z.B. 75% einstellst und eigentlich 10 IE bolen solltest, wird der Bolus-Rechner nur einen Mahlzeitenbolus von 7,5 IE vorschlagen.
Aktivierung des Superbolus im Bolus-Rechner. (Das ist etwas anderes als ein Super Micro Bolus!): Verwende den Superbolus mit Vorsicht und vor allem nicht, bevor Du wirklich verstanden hast, wie er funktioniert. Im Wesentlichen wird das Basalinsulin der nächsten zwei Stunden zum Bolus hinzugefügt und die Basalrate für zwei Stunden auf null gesetzt. AAPS Loop-Funktionen werden deaktiviert - also mit Vorsicht verwenden! If you use SMB AAPS looping functions will be disabled according to your settings in „Max minutes of basal to limit SMB to“, if you do not use SMB looping functions will be disabled for two hours. Details on super bolus can be found here.
Aktionen
A tab offering multiple buttons to take actions in AAPS.
Automatisierung
A tab for managing your Automations, starting at Objective 10.
SMS Kommunikator
Allows remote caregivers to control some AAPS features via SMS, see SMS Commands for more setup information.
Essen
Zeigt die im Nahrungsmittel-Editor erfassten Einträge an. Weitere Informationen zur Einrichtung der Nahrungsmitteldatenbank findest Du im Nightscout Readme.
Note: Entries cannot be used in the AAPS calculator. (reine Anzeigefunktion)
Wear
Monitor and control AAPS using your Android Wear watch (see page Watchfaces). In den Einstellungen (Zahnradsymbol) kannst Du die Variablen festlegen, die bei der Berechnung eines über die Uhr gegebenen Bolus berücksichtigt werden sollen (z.B. 15‘-Trend, COB…).
Wenn Du Deinen Loop von der Uhr aus steuern willst (z.B. Bolus abgeben), aktiviere „Steuerung durch die Uhr“.
Über den Wear Tab oder das Hamburger Menü (oben links, falls der Wear Tab nicht angezeigt wird) kannst Du
Alle Daten erneut senden. Dies kann hilfreich sein, wenn die Uhr längere Zeit außer Reichweite war und du die Informationen an die Uhr pushen willst.
Über das Smartphone die Einstellungen auf der Uhr öffnen.
Wartung
Access this tab to export / import settings.
Konfigurations-Generator
This current tab.